Accu-Chek Go Manual de servicio Pagina 155

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 208
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 154
155
Measurement and storage conditions
12
12 Measurement and storage conditions
Pay close attention to the following points. They affect the dependability
of your meter as well as the accuracy of your results.
A meter that has been dropped can produce false results.
12.1 Temperature range
O Without the battery the meter may be kept at between -25 °C and
+70 °C.
O With the battery the meter may be kept at between -10 °C and
+50 °C.
At temperatures above +50 °C the battery could leak and
damage the meter.
At temperatures below -10 °C the battery does not have
enough power to keep the internal clock functioning.
O For blood glucose tests and performance checks the permitted
range is from +10 °C to +40 °C.
Vista de pagina 154
1 2 ... 150 151 152 153 154 155 156 157 158 159 160 ... 207 208

Comentarios a estos manuales

Sin comentarios