130
Symbols, display messages and troubleshooting
131
Symbols, display messages and troubleshooting
9
This message/symbol
appears…
and means:
The bar code information from the
test strip drum has been lost.
Press the Test button to turn the
meter off. Open and close the drum
compartment.
The meter then rotates the drum
and reads the bar code. The drum
symbol and the bottle symbol flash
in the display when the next test is
performed. Perform a quality control
check. You may use up all the test
strips remaining in the drum. The
90-day use-by period will no longer
be properly calculated for this drum.
The drum symbol continues to flash
in the display until you have inserted
a new drum.
The bar code on the test strip drum
is damaged.
This message/symbol
appears…
and means:
(cont.) If you inserted the first drum from
a pack, remove the drum from the
meter and return it to its container.
Insert the second drum from the
same pack in the meter and use
all 17 test strips. When it is time to
replace the drum, insert the first
drum with the damaged bar code.
When the error message is displayed
again, press the S button. The meter
uses the bar code from the preced-
ing drum. The drum and bottle sym-
bols flash in the display. The meter
requests you to perform a quality
control check. You may use all 17
test strips for testing. The drum sym-
bol continues to flash in the display
until you have inserted a new drum.