BLOOD GLUCOSE METER / GLYCOMÈTREUser’s ManualManuel de l'utilisateur

Pagina 2

1. Make sure the meter is off.2. Turn the meter over. 3. Remove the old code chip (if there is one in the meter) and discard it.Coding the Meter▲!C

Pagina 3

4. Turn the code chip over so the code number faces away from you. Push it into the code chip slot until it stops.5. Leave the code chip in the mete

Pagina 4 - Aviva System

Adjusting the Time and Date—First Time UseSetting the correct time and date in the meter is important if you use the meter memory or if you want to do

Pagina 5 - Need Help?

4. Press and release or to adjust the minutes. Press to set the minutes.5. Repeat to set the am/pm, month, day, and year.6. Once you set the year

Pagina 6 - Contents

Using the Accu-Chek Aviva System• Only use Accu-Chek Aviva Test Strips.• Change the code chip every time you open a new box of test strips. • S

Pagina 7

Chapter 2: Testing Your Blood GlucosePerforming a Blood Glucose TestBefore you perform your first blood glucose test, set up the meter correctly. You

Pagina 8 - The Accu-Chek Aviva Meter

Performing a Blood Glucose Test (continued)4. Make sure the code number on the display matches the code number on the test strip container. If you mi

Pagina 9

6. Poke your fingertip with the lancing device.7. Gently squeeze your finger to assist the flow of blood. This helps you get a blood drop.15

Pagina 10 - Coding the Meter

Performing a Blood Glucose Test (continued)8. Touch the blood drop to the front edge of the yellow window of the test strip. Do not put blood on top

Pagina 11

Note: After a successful test, the meter turns itself off five seconds after the test strip is removed.9. The result appears on the display. If you w

Pagina 13

Flagging Test Results If you wish, you can “flag” your test result with an asterisk ( ) to mark a special result. For example, you might want to flag

Pagina 14

Alternate Site Testing (AST)You have the option of testing other sites on your body besides the fingertip. Blood obtained from a fingertip can be used

Pagina 15

Do not change your treatment because of just one result. NEVER ignore symptoms of high or low blood glucose. If your blood glucose resu

Pagina 16

You need the meter, a test strip, a lancing device designed for AST, and a lancet.1. Prepare the lancing device.2. Insert the test strip into the met

Pagina 17

4. A test strip and flashing blood drop symbol appear on the display.5. Press the lancing device firmly against a fleshy area on the alternate site.

Pagina 18

6. Trigger the lancing device while keeping steady pressure on the site. Apply pressure to the site with the lancing device to assist the flow of blo

Pagina 19

24Normal Blood Glucose ValuesThese test strips are calibrated to deliver plasma-like results. The normal fasting blood glucose range for an adult with

Pagina 20 - Flagging Test Results

25Troubleshooting Checks1. Check if the test strips were expired.2. Check if the cap on the test strip container was always closed tightly.3. Check

Pagina 21 - Alternate Site Testing (AST)

26Troubleshooting Checks5. Check if you followed the testing steps.6. Check if the code number on the meter display matches the code number on the

Pagina 22

27Symptoms of High or Low Blood GlucoseBeing aware of the symptoms of high or low blood glucose can help you understand your test results and decide w

Pagina 23

Whether the Accu-Chek Aviva Meter is your first blood glucose meter or you have used a meter for some time; please take the time to read this manual c

Pagina 24

28Chapter 3: Meter Memory, Set-up, and DownloadingMemoryStoring Test ResultsThe meter automatically stores up to 500 blood glucose test results with t

Pagina 25

29Notes: • Do not change your therapy based on one individual result in memory.• The memory is not lost when you replace the battery. You do need to

Pagina 26 - Normal Blood Glucose Values

30Viewing Test ResultsPress either or to view the memory. The most recent result appears. Press to view previous results in order.ORPress to l

Pagina 27 - Unusual Test Results

31Meter Set-upUsing the Set-up ModeBy using the set-up mode, you can personalize the meter to suit your lifestyle. Here are the features you can custo

Pagina 28

32TIME / DATE(hrs, min, am/pm, month, day, year) BEEPER (on/off)ALARM CLOCK (A-1, A-2, A-3, A-4)HYPO ALARM(off, on, level)You are hereSetting the Time

Pagina 29

334. Press and release to set the hour. The minutes flash.5. Press and release or to adjust the minutes. Press to set the minutes.Note: Pres

Pagina 30 - Storing Test Results

34Setting the Time and Date (continued)6. Repeat to set the am/pm, month, day, and year.7. If you want to set up more options, press and release . If

Pagina 31

35TIME / DATE(hrs, min, am/pm, month, day, year) BEEPER (on/off)ALARM CLOCK (A-1, A-2, A-3, A-4)HYPO ALARM(off, on, level)You are hereSetting the Beep

Pagina 32 - Viewing Test Results

361. Press to turn the meter on. The flashing test strip symbol appears.2. Enter the set-up mode (press and hold for about four seconds). “Set-u

Pagina 33 - Meter Set-up

374. Press or to switch between “On” or “OFF.”5. If you want to set up more options, press and release . If you want to exit, press and hold

Pagina 34 - Setting the Time and Date

▲!2The Accu-Chek® Aviva SystemThe Accu-Chek Aviva Meter is for quantitative blood glucose testing using Accu-Chek Aviva Test Strips. The Accu-Chek Avi

Pagina 35

TIME / DATE(hrs, min, am/pm, month, day, year) BEEPER (on/off)ALARM CLOCK (A-1, A-2, A-3, A-4)HYPO ALARM(off, on, level)You are hereSetting the Alarm

Pagina 36

39When you are setting the time for the alarm clock function, the bell symbol remains on the display and “set-up” flashes continuously.1. Press to

Pagina 37 - Setting the Beeper On/Off

40Setting the Alarm Clock Function (continued)4. Press and release or to switch between “On” or “OFF.” Press and release to set your choice.5. I

Pagina 38

417. The minutes flash. Press and release or to select 00, 15, 30, or 45. These are the only choices.8. Press and release to set the minutes.

Pagina 39

42TIME / DATE(hrs, min, am/pm, month, day, year) BEEPER (on/off)ALARM CLOCK (A-1, A-2, A-3, A-4)HYPO ALARM(off, on, level)You are hereSetting the Hypo

Pagina 40

432. Enter the set-up mode (press and hold for about four seconds). “Set-up” and the hour flash on the display.3. Press and release repeatedly u

Pagina 41

44Setting the Hypoglycemic (Hypo) Alarm Function (continued)5. If you select “On,” “set-up” and flash. The display shows 3.9 mmol/L.6. Press and r

Pagina 42

Downloading Your Results to a Computer or PDAYou can transfer your stored results to a computer to track, identify patterns, and print. Transferring

Pagina 43

466. Put the meter on a flat surface. Point the two IR windows toward each other. They should be 3 to 10 cm apart.7. Do not move the infrared cable

Pagina 44

47Chapter 4: Control TestingWhy Perform Control TestsPerforming a control test lets you know that the meter and test strips are working properly to gi

Pagina 45

Need Help?Call Accu-Chek Customer Care at 1-800-363-7949.Our address: Roche Diagnostics 201, Boulevard Armand Frappier Laval, Québec (CANADA) H7V 4A2

Pagina 46

48About the Control Solutions• Only use Accu-Chek Aviva Control Solutions.• The meter automatically recognizes the Accu-Chek Aviva Control Solution.

Pagina 47

49Performing a Control TestYou need the meter, a test strip, and control solution Level 1 or Level 2. The control level is printed on the bottle label

Pagina 48

503. Select the control solution you want to test. You will enter the level later in the test.4. Put the meter on a flat surface, like a table.5. Remo

Pagina 49 - Chapter 4: Control Testing

516. Squeeze the bottle until a tiny drop forms at the tip. Touch the drop to the front edge of the yellow window of the test strip. When you see fl

Pagina 50 - About the Control Solutions

52Performing a Control Test (continued)8. Press to set the control level in the meter.9. “OK” and the control result alternate on the display if t

Pagina 51 - Performing a Control Test

53Range (mmol/L)Level 1 1.4–3.1Level 2 14.2–19.1Understanding Control Test ResultsThe label on the test strip container shows the acceptable ranges

Pagina 52

54If the control result is not inside the acceptable range, here are some things you can do to solve the problem:Troubleshooting Checks1. Check if th

Pagina 53

55Troubleshooting Checks6. Check if you followed the testing steps.7. Check if you chose the correct control solution level, either 1 or 2, when you

Pagina 54

Chapter 5: Maintenance and TroubleshootingChanging the Battery1. Open the battery door on the back of the meter by pushing the tab in the direction o

Pagina 55

57Notes:• The meter uses one 3-volt lithium battery, type 2032. This type of battery can be found in many stores. It is a good idea to have a spare b

Pagina 56

ContentsChapter 1: Understanding Your New System ...6 The Accu-Chek Aviva Meter ...

Pagina 57

58Cleaning the MeterCaring for the Accu-Chek Aviva Meter is easy—just keep it free of dust. If you need to clean it, follow these guidelines carefully

Pagina 58 - Changing the Battery

59Maintenance and TroubleshootingThe meter needs little or no maintenance with normal use. It automatically tests its own systems every time you turn

Pagina 59

60Screen Messages and Troubleshooting▲!Never make treatment decisions based on an error message. If you have any concerns, call Accu-Chek Customer Car

Pagina 60 - Cleaning the Meter

61Battery power is low. Change the battery soon. See Chapter 5 “Changing the Battery.”The meter is in set-up mode, waiting for you to change or confir

Pagina 61

62Blood glucose may be higher than the measuring range of the system.Blood glucose may be lower than the measuring range of the system.The test result

Pagina 62

63The test strips expire at the end of the current month. Before the end of the month, insert a new code chip from a new box of test strips and ensure

Pagina 63

64Your blood glucose level may be extremely low, or the test strip may be damaged or not properly inserted. If you see this error message after you ap

Pagina 64

65Not enough blood or control solution was drawn into the test strip for measurement or was applied after the test had started. Discard the test strip

Pagina 65

66Note: If you see any other error screen, please call Accu-Chek Customer Care at 1-800-363-7949.The temperature is above or below the proper range f

Pagina 66

67Chapter 6: Technical InformationProduct LimitationsPlease read the literature packaged with your test strips to find the latest information on produ

Pagina 67

Chapter 4: Control Testing ...47 Why Perform Control Tests ...

Pagina 68

68 Memory capacity 500 blood glucose and 20 control results with time and date Automatic power off 2 minutes Power supply One 3-volt lithium

Pagina 69 - Specifications

69Performance AnalysisThe performance data for the Accu-Chek Aviva System (Accu-Chek Aviva Meter with Accu-Chek Aviva Test Strips) were obtained using

Pagina 70

70Product Safety InformationStrong electromagnetic fields may interfere with the proper operation of the meter. Do not use this meter close to sources

Pagina 71

Consult the instructions for use Caution (refer to accompanying documents). Please refer to safety-related notes in the manual accompanying thi

Pagina 72 - Disposing of the Meter

72Warranty and ServiceRoche Diagnostics warrants the meter against any material and manufacturing defects that may arise during proper use of the inst

Pagina 73 - Explanation of Symbols

73Additional SuppliesThe following supplies and accessories are available from your authorized Roche Diagnostics Diabetes Healthcare Center, pharmacie

Pagina 74 - Warranty and Service

74▲!Information for Healthcare ProfessionalsHealthcare Professionals: Follow the infection control procedures appropriate for your facility.WARNING: P

Pagina 75 - References

75Indexalarm clock function, setting, 38alternate site testing, 19, 74appetite, increased, 27battery, changing, 56battery, installing, 56battery,

Pagina 76

76test results, unusual, 25test strips, Accu-Chek Aviva, 7, 12thirst, increased, 27time and date, setting, 32tingling, 27trembling, 27troubleshooting

Pagina 77

Chapter 1: Understanding Your New SystemThe Accu-Chek Aviva MeterInfrared (IR) Window – Transfers data from the meter to a computer or PDA.Display – S

Pagina 78

Code Chip(for example)Battery – Insert with (+) symbol facing up.Test Strip Container Control Solution BottleGolden End – Insert this end of the test

Comentarios a estos manuales

Sin comentarios