Accu-Chek 98/79/EC Manual de servicio Pagina 12

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 107
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 11
22
Initial steps before testing
23
Initial steps before testing
2
If you continue to hold down the M button, you then see the following
display after approximately 2 seconds.
mmol/L meter mg/dL meter
or
When you release the M button, your meter is in memory mode. As long
as no results have yet been saved, three bars are displayed.
 Press the Test button to turn the meter off.
Notes
i
If additional lines appear during the display test or parts of the
numeric field or of a symbol are missing, please contact the
Accu-Chek Customer Careline free on 0800 701000 (UK) or
1800 709600 (Ireland). If the numeric field

(mmol/L meter)
or

(mg/dL meter) is affected, the meter may not show your
test results correctly.
i
The meter automatically turns off after 60 seconds even if you
keep the M button pressed.
i
You can perform a display check whenever you wish. If there are
results already saved in memory, the most recently saved result
is displayed when you release the M button.
Vista de pagina 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 106 107

Comentarios a estos manuales

Sin comentarios