7.6.1 Program a temporary basal rate
FC_temporary_BR_EN.eps
precentage
temporary
basal rate
RUN
screen
duration
Make sure your insulin pump is in RUN.
Press d to move to the TEMPORARY BASAL
RATE (TBR) menu.
Select with f. The TBR PERCENTAGE screen
appears.
Note If no temporary basal rate is currently active, the percentage
is set to 100%. If a temporary basal rate is active, its
duration and percentage appear on the RUN screen.
Press a to increase or s to decrease
the basal rate.
Note When you increase (or decrease) the temporary basal rate
the duration of your previous temporary basal rate increase
(or decrease) appears automatically, or when using your
insulin pump for the first time, a default value appears.
Comentarios a estos manuales