Data transfer
143
Check the configuration on your insulin pump to
make sure that all parameters are written
correctly.
Set up your insulin pump with a cartridge,
adapter and a new infusion set and put it into
RUN, if necessary.
Note Your insulin pump automatically returns to STOP 15 minutes
h after the last data transfer respectively when the DATA
TRANSFER screen has been selected.
You may press d+a simultaneously to exit while data is
being transferred. The data transfer is interrupted and an
error E12: DATA INTERRUPTED occurs.
See section 10.2.11 “Error E12: DATA INTERRUPTED” for
further information.